Re: Cal-Kit Standards' Definitions


Dr. David Kirkby from Kirkby Microwave Ltd
 

On Sun, 6 Oct 2019 at 19:26, Oristo <ormpoa@...> wrote:

If the NanoVNA could provide a negative Electrical delay it could be
fixed.

IMHO, we really need a way of *properly* defining a calibration kit in
*firmware*. That means as a minimum a delay, (which can be negative), and
at least fixed capacitance C0 to support *professional* cal kits.

To support poorer kits C0, C1, C2, C3, L0, L1, L2, loss terms and
impedance.

Dave.


--