In the NanoVNA-saver the calibration settings, when doing a female
calibration with the supplied NanoVNA calibration kit, where the male SOL
parts are fitted to the FF adaptor, need some special settings until Rune
gets fixing the following:
The Open delay does not work and substituted by a C0 setting as shown in
attached image of 1022.3 fF Do not use the open calibration standard just
leave the FF adaptor open
As Shunt C for the load not yet implemented an inductance og 205pH
introduced as temporary fix
The data are for the NanoVNA-H delivered after end October. The earlier
delivered kits are for Short and Load identical but the delay of FF adaptor
is a bit longer and 70.3ps
Remember the NanVNA internal C0 calibration must be done prior to using the
NanoVNA-saver with the same setup as when calibration with the NanoVNA-saver
Kind regards
Kurt