Thanks - add it to the console command doc.
...Larry
toggle quoted message
Show quoted text
On Friday, October 4, 2019, 12:58:16 p.m. GMT-4, erik@... <erik@...> wrote:
The current firmware allows dumping the internal calibration table with the data command
data [0-6]
where
0: S11
1: S21
and
2 /* error term directivity */
3 /* error term source match */
4 /* error term refrection tracking */
5 /* error term transmission tracking */
6 /* error term isolation */
but it is currently NOT possible to upload calibration tables into the nanoVNA