Re: NanoVNA Saver


 

Hi Jim,
I've studied the file, and it seems the same values were used for the
different calibration standards (OSL as well as I&T).

In a fit of poor design philosophy, I initially elected *not* to let the
buttons in the calibration screen run the sweep for you, but to expect the
user to do this manually for each standard before pressing the respective
button. It's become blatantly clear that this is completely unintuitive,
and I'll be looking at changing it - but for now, when you want to use
in-app calibration, make sure that you run a sweep for each standard, and
after each, press the relevant button in the calibration window. That's
currently the only way to get a valid in-app calibration.

I'll make a better interface for it, but it probably won't be for a couple
of releases (ie. not this coming week ;-)

--
Rune / 5Q5R

On Sat, 21 Sep 2019 at 22:33, <jimcking@...> wrote:

Hi Rune,

Attached is the calibration File.

Thanks Again.

Jim K. K8SLC