Hello Ed,

Op ma 18 jul. 2022 om 08:26 schreef Ed G8FAX <ed@...>:


For DM, the DP paper makes use of two baluns.

Like the HAL document, Fig 16b.

Thus, I see that the ¡®open¡¯ measurement of the DP paper (fig 2 (c) is part
of a series of tests of a particular method for the purposes of modelling,
not an alternative to more conventional means of establishing performance
of a cmc.

But if you have the values you can also calculate the CM rejection (CMR).
But at the end I think that these two articles are more power-supply Choke
and not Transmission Line Choke. Which has a slightly different set of
parameters). The power supply choke have a fCO1 and fCO2, while I think
that the Transmission line Chokes have only fCO1:



Just as are the additional configurations of the HAL document Fig 3 (c),
(d) & (e) with T4 being equivalent to DP paper ¡®open¡¯.
DP Open is an OC S21 measurement. While HAL Fig 3 (c) T2, (d) T3 & (e) T4
are reflection measurements (to determine 'individual' choke parameters).



Interesting to note that in the HAL document, the conclusion is that
results not only depend on components (tolerances?) but also on respective
positioning (lay-out).This reminds me of a quotation I heard many years ago
(can¡¯t remember where/who it came from) =¡±Theoretically, practice and
theory are the same, in practice, they are not!¡±
Certainly for a Choke it is important what is around it. That is why I test
on a styrofoam platform. So it is influenced by other objects (including
hands, etc.:


I have also been refreshing and improving my knowledge on transmission line
transformers (TLT¡¯s) and found for me a very useful reference = Ruthroff
transmission line transformers, Guanella baluns, lowpass and bandpass
filters, and balanced transmission lines using multilayer technology, Riaz
R Sobrany, M.Eng.



Looks an interesting link as it handles 4:1 Ruthroff and Guanella (I think
he only discusses 4:1 and 16:1 Guanella) Thanks. Will study it

¡°Connecting two identical baluns with unknown performance back to back is a
very bad test procedure to use.
All that can be determined is the degree of mismatch at the unbalanced
port and the insertion loss of the balun. This test procedure is incapable
of revealing values for the mode conversion parameters which means that the
CMRR cannot be calculated, and even worse, provides no conclusive evidence
that the balun actually functions as a balun at all!
A procedure to test a balun using a conventional 2-port VNA which
determines the CMMR but does not measure the insertion loss or the port
reflection coefficients is detailed in [113]. This test procedure requires
three precision resistors and the accuracy of the measurement is dependent
upon the accuracy of these resistors.¡±
Indeed one needs to know how good these Baluns are. But one needs something
to go from unbalanced the VNA port to balanced (the DUT).
I was wondering: if one includes the baluns in the calibration, would that
not calibrate that setup?


With ref 113 being ¡°O. M. Woodward, ¡°Balance Quality Measurements on
Baluns,¡± IEEE Transactions on Microwave Theoiy and Techniques, vol. 31, no.
10, pp. 821-824, Oct. 1983.¡± I have not located this, so no further comment
on the technique at present.
Indeed that is the reference I would like to read also. Does someone have a
copy?

I did a little research in to commercial baluns, ¡°mini-Circuits¡± produce a
very wide range of rf transformers (500+), there may be something suitable
there, I have yet to explore possibilities

I think the 1:1:1 T-622-KK81 can do the function.

All the best,

Victor


Re: checking inductance

 

On 7/18/22 10:10 AM, Marc et Nicole Feuggelen-Verbeck wrote:
What you say is correct but that is nitpicking : The value of an inductor does not change with increasing or decreasing frequency . 100?H now equals 100?H just as 100KOhm is equal to 100KOhm .
If not , the laws of physics must be rewritten . F=1/2(pi)(sqrt(LC))
An inductor is *a passive electronic component that stores energy in the form of a magnetic field*. In its simplest form, an inductor consists of a wire loop or coil. The inductance is directly proportional to the number of turns in the coil , the used materials and the envirement but not the frequency.
ideal inductors have zero diameter conductors with zero resistance, and the inductance is
proportional (approximately) to *square* of turns, (if all the turns are equally well magnetically coupled)


The *inductance* of a real inductor most certainly changes with frequency (as captured in standard works like those of Rosa and Grover)


for a variety of reasons - mostly that with finite sized conductors, the current is not distributed evenly, and that distribution depends on the frequency.

There is also an issue of "effective impedance" in that inductance can be partly cancelled by capacitance (in the worst case it completely cancels at self resonance).





End of story
Marco
Op 18/07/2022 om 15:42 schreef Jim Lux:
On 7/18/22 2:20 AM, Marc et Nicole Feuggelen-Verbeck wrote:
The value of an inductor is a given and depends on several factors, but CERTAINLY NOT on the frequency
An ideal inductor, perhaps, but any sort of practical real inductor with parasitic C between turns and the surroundings varies with frequency. The loss also changes with frequency due to skin depth changes (although that's a small effect).
There's also a small inductance change effect due to different distribution of current in windings due to skin effect.

Finally, if the inductor is on a core, pretty much every core material has frequency dependent properties.







Re: checking inductance

 

Marco is correct. The inductance L does not change with frequency (except
for the small non-linear effects of the core materials).

HOWEVER, all real inductors also have parasitic capacitance and resistance,
which causes the Z (impedance, not inductance) to change with frequency.
And these effects are real, and can be quite large. So when the VNA
measures an inductor across a frequency range, you will see the impedance
(not inductance) change from something close to the pure 'L' inductance at
very low frequencies, to where the parasitic capacitance completely cancels
out that inductance at higher frequencies (at the LC resonant frequency)
and then the C becomes dominant at even higher frequencies. So it is often
useful to measure this impedance (or apparent inductance) at the frequency
the inductor will be used - where the modeling of its required valued was
done.

At DC and low frequencies, the Z of an inductor is dominated by its
inductance, and the Z and L are essentially the same. But at higher
frequencies the Z will be dramatically different, and depends highly on the
materials and construction techniques used in the inductor.

The nanovna and other vna's (and/or the software used to control them)
calculate the Z from the measured S11 or S21 parameters, often with a good
degree of accuracy. Then they calculate the L from the Z. There are more
limitations to the accuracy of this calculation.

A good LCR meter is likely a better tool to measure the 'L' of an inductor.

Stan

On Mon, Jul 18, 2022 at 10:11 AM Marc et Nicole Feuggelen-Verbeck <
f8voa54@...> wrote:

What you say is correct but that is nitpicking : The value of an
inductor does not change with increasing or decreasing frequency . 100?H
now equals 100?H just as 100KOhm is equal to 100KOhm .

If not , the laws of physics must be rewritten . F=1/2(pi)(sqrt(LC))

An inductor is *a passive electronic component that stores energy in the
form of a magnetic field*. In its simplest form, an inductor consists
of a wire loop or coil. The inductance is directly proportional to the
number of turns in the coil , the used materials and the envirement but
not the frequency.

End of story

Marco


Op 18/07/2022 om 15:42 schreef Jim Lux:
On 7/18/22 2:20 AM, Marc et Nicole Feuggelen-Verbeck wrote:
The value of an inductor is a given and depends on several factors,
but CERTAINLY NOT on the frequency
An ideal inductor, perhaps, but any sort of practical real inductor
with parasitic C between turns and the surroundings varies with
frequency. The loss also changes with frequency due to skin depth
changes (although that's a small effect).
There's also a small inductance change effect due to different
distribution of current in windings due to skin effect.

Finally, if the inductor is on a core, pretty much every core material
has frequency dependent properties.






--
If You are not part of the solution , then You are the problem <<<





Re: checking inductance

 

What you say is correct but that is nitpicking : The value of an inductor does not change with increasing or decreasing frequency . 100?H now equals 100?H just as 100KOhm is equal to 100KOhm .

If not , the laws of physics must be rewritten . F=1/2(pi)(sqrt(LC))

An inductor is *a passive electronic component that stores energy in the form of a magnetic field*. In its simplest form, an inductor consists? of a wire loop or coil. The inductance is directly proportional to the number of turns in the coil , the used materials and the envirement but not the frequency.

End of story

Marco


Op 18/07/2022 om 15:42 schreef Jim Lux:
On 7/18/22 2:20 AM, Marc et Nicole Feuggelen-Verbeck wrote:
The value of an inductor is a given and depends on several factors, but CERTAINLY NOT on the frequency
An ideal inductor, perhaps, but any sort of practical real inductor with parasitic C between turns and the surroundings varies with frequency. The loss also changes with frequency due to skin depth changes (although that's a small effect).
There's also a small inductance change effect due to different distribution of current in windings due to skin effect.

Finally, if the inductor is on a core, pretty much every core material has frequency dependent properties.





--
If You are not part of the solution , then You are the problem <<<


Re: Test an antenna tuning unit input circuit

 

On 7/18/22 8:24 AM, Observer wrote:
I inserted my nanoVNA, ( fully calibrated as swr ) into the input of my ATU100 auto tuner, with power off.
I got anything but an swr reading. Am I doing it right ?
Before test, I inserted the 50 ohm load at CH0 , and it showed 49.70 ohms 1:1.0025 . So, I assume, my vna is calibrated ok
Some ATUs "remember" the tuning setting when power is off, others don't. (i.e. some use latching relays, others don't).

What you should see, looking into the tuner, is the S11 with a dip at the tuned frequency, and not other places. As you change the tuner's Ls and Cs, the dip moves both in depth and frequency.


In general, the NanoVNA doesn't put out enough power for an auto tuner to actually do the tune cycle. If you have a tuner with a serial remote control (or equivalent) like the AT200PC, you can send inductor up/down and capacitor up/down commands and run a tuning algorithm on a PC that looks at the NanoVNA's measurements.

This is non-trivial.


Re: LogMag and SWR

 

Thanks, you are right. My fingers were writing in dbW but my mind was
working with dBm.

Thanks,
*Clyde K. Spencer*



On Mon, Jul 18, 2022 at 9:58 AM PE0CWK via groups.io <pe0cwk=
[email protected]> wrote:

Hi Clyde,

Reading your message, may I correct your example?
I.m.o. +30 dbW -> 1000 Watt and +10dB*W* -> 10 Watt

-Kees, PE0CWK

Op 18-7-2022 om 14:53 schreef Clyde Spencer:
For reference a LogMag reading of -14dB is the same as a VSWR of 1.5:1.
Any logmag reading less than -14dB is better and better. If you get a
logmag reading of -20dB then you are in very good shape. Several external
things can change the resulting logmag such as ground conductivity,
height
of the horizontal leg of the antenna above ground, or proximity of the
vertical part of the antenna to nearby objects, and the number of ground
radials.

There is one telltale measurement that can point to the difference.
Masure
the feed point impedance. The closer the impedance is to 50 ohms the
lower,
the more negative, the LogMag reading. The impedance will be impacted by
the number of radials and ground conductivity.

The logmag is the return loss in dB. EXAMPLE if the forward power is 100
watts, +30dBw, and the reflected power is +10dB, 1 watt, then the return
loss, i.e logmag, will be -20dB.

Hope this helps,



*Clyde K. Spencer*



On Sun, Jul 17, 2022 at 4:17 PM<cariboome@...> wrote:

Good day. I'm fairly new to both amateur radio and this marvelous
device,
the NanoVNA H4 v4.3 sold by Seesii on Amazon.ca
I find tuning my many inverted V antennas using a manual tuner extremely
effective using this Nanovna -- set to s11 on 3 parameters for logmag,
Smith and SWR traces.

My question is: for some tuning, I often show a very low SWR (<1:1.05)
with a logmag reading of -15 or thereabouts, and another tune showing
about
the same SWR but a logmag of -40, -50 or even -90 dB. The logmag trace
sometimes appearing as a very steep and deep notch, often at or in the
SWR
trace and other times just a shallow dip. The Smith typically shows a
fairly consistent 50 ohms plus or minus 1 or 2 at tuned resonance. Is
this
a representation of reflected loss and VSWR being shown real time side
by
side? Is it better to have a logmag reading of -90 dB let's say, as
well as
a very low SWR -- vs very low SWR and a "higher" logmag of -20dB for
instance?

I sincerely hope I'm expressing myself sensibly...
--
- Dave
-VA7WNW-













Re: LogMag and SWR

 

Thank you all. I've attached images of some early readings made with the NanoVNA as I explored. It's an extremely valuable tool. I'm pleased to confirm I am using it correctly. The traces I was getting looked, to my ignorant eyes, too good to be true! But it is true: my first two attempts at antenna making are wildly successful!

--
-VA7WNW-


Re: LogMag and SWR

 

Hi Clyde,

Reading your message, may I correct your example?
I.m.o. +30 dbW -> 1000 Watt and +10dB*W* -> 10 Watt

-Kees, PE0CWK

Op 18-7-2022 om 14:53 schreef Clyde Spencer:

For reference a LogMag reading of -14dB is the same as a VSWR of 1.5:1.
Any logmag reading less than -14dB is better and better. If you get a
logmag reading of -20dB then you are in very good shape. Several external
things can change the resulting logmag such as ground conductivity, height
of the horizontal leg of the antenna above ground, or proximity of the
vertical part of the antenna to nearby objects, and the number of ground
radials.

There is one telltale measurement that can point to the difference. Masure
the feed point impedance. The closer the impedance is to 50 ohms the lower,
the more negative, the LogMag reading. The impedance will be impacted by
the number of radials and ground conductivity.

The logmag is the return loss in dB. EXAMPLE if the forward power is 100
watts, +30dBw, and the reflected power is +10dB, 1 watt, then the return
loss, i.e logmag, will be -20dB.

Hope this helps,



*Clyde K. Spencer*



On Sun, Jul 17, 2022 at 4:17 PM<cariboome@...> wrote:

Good day. I'm fairly new to both amateur radio and this marvelous device,
the NanoVNA H4 v4.3 sold by Seesii on Amazon.ca
I find tuning my many inverted V antennas using a manual tuner extremely
effective using this Nanovna -- set to s11 on 3 parameters for logmag,
Smith and SWR traces.

My question is: for some tuning, I often show a very low SWR (<1:1.05)
with a logmag reading of -15 or thereabouts, and another tune showing about
the same SWR but a logmag of -40, -50 or even -90 dB. The logmag trace
sometimes appearing as a very steep and deep notch, often at or in the SWR
trace and other times just a shallow dip. The Smith typically shows a
fairly consistent 50 ohms plus or minus 1 or 2 at tuned resonance. Is this
a representation of reflected loss and VSWR being shown real time side by
side? Is it better to have a logmag reading of -90 dB let's say, as well as
a very low SWR -- vs very low SWR and a "higher" logmag of -20dB for
instance?

I sincerely hope I'm expressing myself sensibly...
--
- Dave
-VA7WNW-








Re: Using VNA to test NFC tags

 

On 7/18/22 5:10 AM, Larry Martin wrote:
Jim Lux wrote:
The actual load may or may not be 50 ohms in the design circuit.
Hi Jim, thanks for a thoughtful response.? But I'm not sure transmission line effects play into this test at 13.56 MHz, where wavelength/10 is around 10 feet.? They could explain some things I see at UHF (915 MHz, wavelength/10 more like an inch).? There are fewer than 50 part numbers of HF/NFC RFID ICs in common use.? They are not specified by impedance but by capacitance, which must be balanced against their respective coils.? Most capacitance specs cluster around 15 pF or 50 pF, but the "50" pF rating does not relate to "50" ohm impedance.
Can you give a sample part #, and I'll go look at the data sheet.

Further, if we had to worry about tags' internal impedance, some tags would not work with some readers.? HF/NFC tags work with "any" HF RFID reader, not just ones that match the impedance or capacitance of the inlay design.? In this test, the NanoVNA CH0 stands in for the RFID reader, and should have a similar level of cross operability.

In a real reader, the requirement for the transmitter is "produce at least X magnetic field" and for the receiver "sense no less than Y magnetic field".

For a active reader passive tag, don't the tags just rectify the transmitted signal to produce DC to run the internal tag chip?



To me, the LOGMAG curves in the linked image () represent CH0 transmit power that is not reflected back to CH0 because it is coupled to the RFID tag and consumed by the RFID chip.? I'm trying to see what-all can be deduced from that measurement.
That's only sort of true. The VNA has a 50 ohm source impedance, so you'd need to take that into account.

Like I said, model it as a non-ideal transfomer (with some TBD turns ratio) that transforms some unknown load impedance which is a resonator with loss.


Re: checking inductance

 

On 7/18/22 2:20 AM, Marc et Nicole Feuggelen-Verbeck wrote:
The value of an inductor is a given and depends on several factors, but CERTAINLY NOT on the frequency
An ideal inductor, perhaps, but any sort of practical real inductor with parasitic C between turns and the surroundings varies with frequency. The loss also changes with frequency due to skin depth changes (although that's a small effect).
There's also a small inductance change effect due to different distribution of current in windings due to skin effect.

Finally, if the inductor is on a core, pretty much every core material has frequency dependent properties.


Re: Common ground and 2-port measuring

 

On 7/17/22 11:26 PM, Ed G8FAX wrote:
Hi Victor,
¡°Connecting two identical baluns with unknown performance back to back is a very bad test procedure to use.
All that can be determined is the degree of mismatch at the unbalanced port and the insertion loss of the balun. This test procedure is incapable of revealing values for the mode conversion parameters which means that the CMRR cannot be calculated, and even worse, provides no conclusive evidence that the balun actually functions as a balun at all!

E.g. a "back to back" test of a pair of coax to banana jack adapter (which is not a balun) would show good S11 and S21.


Re: LogMag and SWR

 

For reference a LogMag reading of -14dB is the same as a VSWR of 1.5:1.
Any logmag reading less than -14dB is better and better. If you get a
logmag reading of -20dB then you are in very good shape. Several external
things can change the resulting logmag such as ground conductivity, height
of the horizontal leg of the antenna above ground, or proximity of the
vertical part of the antenna to nearby objects, and the number of ground
radials.

There is one telltale measurement that can point to the difference. Masure
the feed point impedance. The closer the impedance is to 50 ohms the lower,
the more negative, the LogMag reading. The impedance will be impacted by
the number of radials and ground conductivity.

The logmag is the return loss in dB. EXAMPLE if the forward power is 100
watts, +30dBw, and the reflected power is +10dB, 1 watt, then the return
loss, i.e logmag, will be -20dB.

Hope this helps,



*Clyde K. Spencer*

On Sun, Jul 17, 2022 at 4:17 PM <cariboome@...> wrote:

Good day. I'm fairly new to both amateur radio and this marvelous device,
the NanoVNA H4 v4.3 sold by Seesii on Amazon.ca
I find tuning my many inverted V antennas using a manual tuner extremely
effective using this Nanovna -- set to s11 on 3 parameters for logmag,
Smith and SWR traces.

My question is: for some tuning, I often show a very low SWR (<1:1.05)
with a logmag reading of -15 or thereabouts, and another tune showing about
the same SWR but a logmag of -40, -50 or even -90 dB. The logmag trace
sometimes appearing as a very steep and deep notch, often at or in the SWR
trace and other times just a shallow dip. The Smith typically shows a
fairly consistent 50 ohms plus or minus 1 or 2 at tuned resonance. Is this
a representation of reflected loss and VSWR being shown real time side by
side? Is it better to have a logmag reading of -90 dB let's say, as well as
a very low SWR -- vs very low SWR and a "higher" logmag of -20dB for
instance?

I sincerely hope I'm expressing myself sensibly...
--
- Dave
-VA7WNW-






Re: checking inductance

 

Inductance is independent of frequency. Inductance only depends on the materials used (this also means that the environment of the element is important). If you vary the frequency, the inductance NEVER changes.

Marco , F8VOA

Op 18/07/2022 om 14:18 schreef Reinier Gerritsen:
Inductance may vary with frequency, e.g. when there is a ferrite core. The core material's properties are frequency dependent. Also, the effect of the interwinding capacitance increases with frequency and that will change the observed inductance.
Reinier

On July 18, 2022 11:20 AM Marc et Nicole Feuggelen-Verbeck <f8voa54@...> wrote:

The value of an inductor is a given and depends on several factors, but
CERTAINLY NOT on the frequency

F8VOA , Marco

Op 18/07/2022 om 09:27 schreef Observer:
Its a fact, that inductance value, vary according to frequency.
How does the nanovna measure ?
Does it use the saved selected frequency range inside the nano ?

--
If You are not part of the solution , then You are the problem <<<


Re: checking inductance

 

Inductance may vary with frequency, e.g. when there is a ferrite core. The core material's properties are frequency dependent. Also, the effect of the interwinding capacitance increases with frequency and that will change the observed inductance.
Reinier

On July 18, 2022 11:20 AM Marc et Nicole Feuggelen-Verbeck <f8voa54@...> wrote:


The value of an inductor is a given and depends on several factors, but
CERTAINLY NOT on the frequency

F8VOA , Marco

Op 18/07/2022 om 09:27 schreef Observer:
Its a fact, that inductance value, vary according to frequency.
How does the nanovna measure ?
Does it use the saved selected frequency range inside the nano ?


Re: Using VNA to test NFC tags

Larry Martin
 

Jim Lux wrote:
The actual load may or may not be 50 ohms in the design circuit.
Hi Jim, thanks for a thoughtful response.? But I'm not sure transmission line effects play into this test at 13.56 MHz, where wavelength/10 is around 10 feet.? They could explain some things I see at UHF (915 MHz, wavelength/10 more like an inch).? There are fewer than 50 part numbers of HF/NFC RFID ICs in common use.? They are not specified by impedance but by capacitance, which must be balanced against their respective coils.? Most capacitance specs cluster around 15 pF or 50 pF, but the "50" pF rating does not relate to "50" ohm impedance.

Further, if we had to worry about tags' internal impedance, some tags would not work with some readers.? HF/NFC tags work with "any" HF RFID reader, not just ones that match the impedance or capacitance of the inlay design.? In this test, the NanoVNA CH0 stands in for the RFID reader, and should have a similar level of cross operability.

To me, the LOGMAG curves in the linked image () represent CH0 transmit power that is not reflected back to CH0 because it is coupled to the RFID tag and consumed by the RFID chip.? I'm trying to see what-all can be deduced from that measurement.

Thanks for weighing in,
Larry Martin


Re: checking inductance

 

The term "inductance" has two meanings. It may mean the inductance exhibited by a coil or circuit at a frequency of interest. It also may mean the inductance at a low frequency where variation with frequency is negligible. The two values can differ due to coil self-resonance, change in effective coil diameter with frequency, change in dielectric permittivity or core permeability with frequency, or other factors. The first value usually is just called inductance. The second may be called true inductance, ideal inductance, or low-frequency inductance. The confusion comes when it too is called inductance.

When referring to inductance in public, you can avoid confusion by stating which meaning you're using.

Brian


Re: checking inductance

 

The NanoVNA will measure inductance over a frequency range, which you
configure in the sweep settings, just like with other measurements.

In theory, the inductance does not depend on the frequency, but the
inductive reactance does.
I'd like someone to correct me if I'm wrong, but from what I can see, VNAs
can't actually directly measure inductance (is there anything that can?),
and instead measures impedance, and then estimates inductance from
impedance. Since everything is an RLC circuit (in coils, don't forget the
resistance of the conductor, and the capacitance between the turns, which
affects the coil's self-resonant frequency), the estimate of the inductance
will change with frequency, sometimes drastically (again, keep the
self-resonant frequency in mind).

§á§à§ß, 18. ?§å§Ý 2022. §å 09:27 Observer <tvstreamdevice@...> ?§Ö
§ß§Ñ§á§Ú§ã§Ñ§à/§Ý§Ñ:

Its a fact, that inductance value, vary according to frequency.
How does the nanovna measure ?
Does it use the saved selected frequency range inside the nano ?






Re: checking inductance

 

The value of an inductor is a given and depends on several factors, but CERTAINLY NOT on the frequency

F8VOA , Marco

Op 18/07/2022 om 09:27 schreef Observer:
Its a fact, that inductance value, vary according to frequency.
How does the nanovna measure ?
Does it use the saved selected frequency range inside the nano ?



--
If You are not part of the solution , then You are the problem <<<


Re: Correction of error introduce by a transmission line connect to the VNA port 1

Diane BONKOUNGOU
 

Hello Roger,
I have two coaxial cables that came with the NanoVNA. I cut one of them to
make the solder. I use the cable I didn't cut to do the calibration and
after connecting the one I cut, I correct the delay. That's why the delay
is so low. When I do the calibration directly at the input of the NanoVNA
and connect my SMA cable, the delay correction is difficult for me, the
phase(pink curve) is triangular like in the attached picture. When I made
the calibration at the cable I didn't cut and I connect the cable I have
cut the phase is just shifted I don't have triangular thing on the curve.
Best regards.



Le ven. 15 juil. 2022 ¨¤ 22:54, Roger Need via groups.io <sailtamarack=
[email protected]> a ¨¦crit :

On Fri, Jul 15, 2022 at 11:44 AM, Diane BONKOUNGOU wrote:


Hello,
Thanks for your response I've got -462ps for the delay correction is it
good?
Best regards.
Your display looks OK but 462 ps would be the value for a very short
transmission line - about 4.5 cm long. What is the length of your
transmission line and what kind of coax is it?

Here is a video you might find interesting. It shows how to measure an
antenna located inside a product...



Roger






checking inductance

 

Its a fact, that inductance value, vary according to frequency.
How does the nanovna measure ?
Does it use the saved selected frequency range inside the nano ?