Keyboard Shortcuts
ctrl + shift + ? :
Show all keyboard shortcuts
ctrl + g :
Navigate to a group
ctrl + shift + f :
Find
ctrl + / :
Quick actions
esc to dismiss
Likes
- Nanovna-Users
- Messages
Search
Re: connector swap
RE: "First BNC up to 3ghz is a reliable connector, no different from TNC.
Caveat; use good stuff and make sure you do not interchange 50 ohm BNC with 75 ohm BNC, really they are different!" Yes they are different, look at the photo at The 75 ohm BNCs are much more delicate and likely to be damaged because they lack the protection of the 50 ohm BNCs However, they are physically inter-matable with no damage. The center pin OD and shield "fongers' ID are identical, so can be inter-mated wtih no damage. The impedance bump is probably not too bad below 5MHZ because before ATSC, when NTSC was used, virtually every BNC connector installed on 75 ohm coax and used in NTSC video equipment was the 50 ohm variety. . |
Re: The T-Check confusion
Hi Kurt.
Thanks. This is the switch I bought: SIVERS LAB PHILIPS COAXIAL SWITCH 28V DC-18Ghz PM7553 Isolation below 1GHz is better than 80dB so I guess below 300MHz good enough. As I am currently the maintainer for the TAPR-VNA software so I can add whatever I want :-) Using the scikit-rf implementation as example I should be able to add (limited) 10/12 term calibration functionally to the TAPR-VNA SW, including controlling the transfer relay. Lately I have been trying to optimize and tune various filters such as a 2GHz cavity filter and as I have to tune 5 resonators it would be nice if I did not need to reverse the DUT all the time as for best tuning you need to see S11,S21,S22 and S12 together. So it is more a matter of being kind to SMA cables and connectors and making tuning faster and easier. For my home build VNA's I can also buy/build a second bridge, add a 4th receiver and put the transfer switch between the LO and the bridges. This will give me a fairly symmetrical setup allowing measurement of the errors in the bridge impedance. Disadvantage is the transfer switch is locked inside one VNA So many options to choose from, sounds like a hobby -- NanoVNA Wiki: /g/nanovna-users/wiki/home NanoVNA Files: /g/nanovna-users/files Erik, PD0EK |
Re: The T-Check confusion
Hi Erik
Using the transfer switch is just a lazy method to swap the direction of the DUT. Actually dependent of the transfer switch quality you bought you make get poorer isolation between port 1 and port2. Often the isolation in the region of 60dB for such transfer switches. Although you have isolation calibration e.g. in forward direction it will not be the same in reverse direction. With the NanoVNA + softwares you cannot do independent calibration in the forward and reverse direction so it is not helping you, on the contrary. Nothing is better than swapping the DUT direction when measuring forward and reverse. You will have different calibration result via the transfer switch for forward and reverse. I did in the past a test set with 4pc. 18GHz SMA microwave relays and used rigid interconnection cable of same phase length. I had isolation up to 140dB with this setup, which was made for my N2PH VNA going to 60MHz. The N2PK VNA was controlled by the VNWA software, which both has separate calibration in forward and reverse direction and did automatic controlled the test set. This way it work also with a transfer relay which I also have works the same way. So bottom line.. You may develop an interface for you homemade VNA controlling the transfer relay and implement the needed error correction in software. There are no free lunches in this game. I have no idea if the TAPR VNA software have such facilities. To the general question about the hobby level then by adding a 10dB attenuator (included in the calibration) between DUT and RX port linearize the measurements quite a lot, so that would be enough in daily use for the accuracy you are after. Kind regards Kurt -----Oprindelig meddelelse----- Fra: [email protected] <[email protected]> P? vegne af erik@... Sendt: 9. december 2019 12:47 Til: [email protected] Emne: Re: [nanovna-users] The T-Check confusion I've found and bought an affordable SMA transfer switch (2x2 switch) on eBay and I was hoping it would allow me to extend nanoVNA or one of my home build VNA's to full two port capability by extending the calibration routines in the TAPR VNA software Do I understand correctly that, as you are not able to measure all errors using a transfer switch connected to port 1 and port 2, you can only do the TwoPortOnePath calibration? What would be the impact of this in practice???? Given I can measure and ensure both ports are rather close to a real 50 ohm using a second VNA: would the inability to correct some errors have a major impact on practical measurements of filters that are sensitive to port mismatch? Remember for me this is all hobby use and being able to measure with an error less then some/one dB is great -- NanoVNA Wiki: /g/nanovna-users/wiki/home NanoVNA Files: /g/nanovna-users/files Erik, PD0EK |
Re: Using the Nano as a receiver?
NanoVNA has receiver, but it cannot digitize receiver output with enough bandwidth to listen AM or SSB radio. You can try to connect your PC soundcard to SA602 outputs and use HDSDR to listen radio, but there is no low pass filter, so you will get a lot of images. And such modification will ruin VNA functions.
|
Re: The T-Check confusion
Hi Erik
Just one more practical comment. Whatever you do the single forward and the single reverse measurement s1p files does not carry any information about the port mismatch to take into account for the T-Check calculation. And besides that I consider this discussion is only of pure interest how T-Check is done the right way, and my report was only to demonstrate some facts. I do not think 99% of the NanoVNA users are benefitted if such facilities was implemented in future and would not try to motivate such a initiative. This product is great as is and a calibration is tested quite well investing in a semirigid cable of some 25cm both for frequency range up to 500MHz and 900MHz. To 1500MHz is does not show useable trace along the Smithchart circumference. Kind regards Kurt -----Oprindelig meddelelse----- Fra: [email protected] <[email protected]> P? vegne af erik@... Sendt: 9. december 2019 08:57 Til: [email protected] Emne: Re: [nanovna-users] The T-Check confusion Kurt, It appears (in my limited understanding) the poor matching to Z0 of CH1 (port 2) is the biggest reason why the T-check is not performing well on the nanoVNA. Would you agree that using a 10dB attenuator between DUT and CH1 (port 2) would solve most of this? What would be the impact of the remaining non-compensated delay? Should we add to the instructions that, for best nanoVNA 2 port DUT measurements one should seriously consider using a 10dB attenuator between DUT and CH1 (port 2) till new HW becomes available that solves the poor matching of CH1 (port2)? -- NanoVNA Wiki: /g/nanovna-users/wiki/home NanoVNA Files: /g/nanovna-users/files Erik, PD0EK |
Re: Using the Nano as a receiver?
Read here:
/g/nanovna-users/topic/34079496#2165 -- NanoVNA Wiki: /g/nanovna-users/wiki/home NanoVNA Files: /g/nanovna-users/files Erik, PD0EK |
Re: The T-Check confusion
I've found and bought an affordable SMA transfer switch (2x2 switch) on eBay and I was hoping it would allow me to extend nanoVNA or one of my home build VNA's to full two port capability by extending the calibration routines in the TAPR VNA software
Do I understand correctly that, as you are not able to measure all errors using a transfer switch connected to port 1 and port 2, you can only do the TwoPortOnePath calibration? What would be the impact of this in practice???? Given I can measure and ensure both ports are rather close to a real 50 ohm using a second VNA: would the inability to correct some errors have a major impact on practical measurements of filters that are sensitive to port mismatch? Remember for me this is all hobby use and being able to measure with an error less then some/one dB is great -- NanoVNA Wiki: /g/nanovna-users/wiki/home NanoVNA Files: /g/nanovna-users/files Erik, PD0EK |
Using the Nano as a receiver?
Hi,
I'm quite new to the topic. I learned a lot about the bad design of commonly available antennas and have modified quite a number since then with the help of the Nano. From what I understand, a VNA is also a receiver (of its own signals). Is there a way to use it like a "scanner"? I do occasionally have to deal with Chinese HF transmitter setups, that only say "UHF" or "VHF" but do not specify the frequency. Normally I go to the company's testlab and have it checked for me. Is there any practical way to use the NanonVNA for that purpose? Thomas |
Re: The T-Check confusion
Hi Gabriel
You are right it is a matter og software and for that matter firmware. As the NanoVNA is presently it does not have facilities embedded for neither calibration kit data and full 10/12 term error correction and none of the present software packages I know off has. That counts for NanoVNA-saver, NanoVNAsharp, NanoVNApartner or the TAPR adjusted program. Thee link you provide is not implemented in the NanoVNA-saver but of course of interest for Rune to consider. You say it is trivial to use in NanoVNA-saver but that requires a "software developer brain" my comment are based on what is on hand for everyone. The VNWA is actually a full two port devise as when fitted with a testset it swaps the direction so your comment is not entirely correct. It has besides a feature in the software (by pressing the F2 key) to do a full two port measurement, without having a testset attached and that is done by mechanically reverting the DUT between the forward and reverse sweep. So you study of the VNWA manual has not been complete. Kid regards Kurt -----Oprindelig meddelelse----- Fra: [email protected] <[email protected]> P? vegne af Gabriel Tenma White Sendt: 9. december 2019 09:06 Til: [email protected] Emne: Re: [nanovna-users] The T-Check confusion On Mon, Dec 9, 2019 at 12:54 AM, Kurt Poulsen wrote: This is wrong. I double checked the VNWA user guide and it is also a T/R VNA just like the Nano, not a full two port VNA. Therefore any calibration types used on the VNWA also apply to the Nano, it's just a matter of software. What you are describing is the Two Port One Path calibration, which is nothing new and there is an implementation of it in scikit-rf: Since scikit-rf is a Python library it is also trivial to use in nanovna-saver. |
Re: errors of "error" models
@ Gary O'Neil, N3GO - 8 December 2019
/g/nanovna-users/message/8024 Dear Gary, Once again, we thank you for your time. You are right. Yes, this is indeed an endless, multidisciplinary, research and one of its objectives is to bridge the gap between the now and the past. Yes, we tried to see the fundamental ideas of quantities and their ratio measurements in our sow, particularly in the two-port network case. Well, this is not at all a simple task as it may sounds, because it has to do with elementary concepts lying on the historical boundaries between science and philosophy. Anyway, after many years of returning attempts, we discovered yesterday that we had to restore the "multiplicative balance" in the two basic, well-known, linear equations, that is instead e.g. of the mathematical: b1=S11a1+S12a2, the "physical" one: 1.b1=S11.a1+S12.a2. That was all. Finally, yes, we also know that you are also absolutely right at all of the rest matters you rise, including the badly shaped o, O, 0, 1 and l of MS Courier New font and especially in this particular size. For that last one, we will try to re-use the font of our choice: Liberation Mono, in an appropriate size. of course. As of the above matters, unfortunately enough, we have to invoke, once again, our permanent excuse : in our sow, this is a research in progress and thus we have to find time to answer all of them. We are terribly sorry for the inconvenience caused. Please, accept our apologies. Best regards, gin&pez@arg |
Re: The T-Check confusion
Hi Eric
A 10dB att in the Ch1 path will not cure the problem as you also have a (invisible) mismatch in the Ch0 sourse impdance as well. So two attenuator might give some further improvements. But it is a fundamental problem as the 10/12 term error correction "totally" remove the tx and rx port mismatch and establish pure 50 ohm source and load impedance in the two calibration planes at the end of the two test cables. Always remember the test cables are an integral part of the VNA instrument and influence removed by the calibration. For the VNWA two custom traces with the expression SS and SL shows the source impedance (SS) and load impedance (SL) after a calibration at the two mentioned calibration planes. However the NanoVNA or for that matter any of the NanoVNA software has such facilities to do a complete S11/S21/S12/S22 measurement controlling a test switch reverting the DUT direction or any other smart tricks for passive bi-direction DUT's Kind regards Kurt -----Oprindelig meddelelse----- Fra: [email protected] <[email protected]> P? vegne af erik@... Sendt: 9. december 2019 08:57 Til: [email protected] Emne: Re: [nanovna-users] The T-Check confusion Kurt, It appears (in my limited understanding) the poor matching to Z0 of CH1 (port 2) is the biggest reason why the T-check is not performing well on the nanoVNA. Would you agree that using a 10dB attenuator between DUT and CH1 (port 2) would solve most of this? What would be the impact of the remaining non-compensated delay? Should we add to the instructions that, for best nanoVNA 2 port DUT measurements one should seriously consider using a 10dB attenuator between DUT and CH1 (port 2) till new HW becomes available that solves the poor matching of CH1 (port2)? -- NanoVNA Wiki: /g/nanovna-users/wiki/home NanoVNA Files: /g/nanovna-users/files Erik, PD0EK |
Re: The T-Check confusion
I've deliberately chosen not to rely on scikit-rf in NanoVNA-Saver. Library
reliance is kept at a minimum, and figuring out the mathematics behind the calibration was part of the fun of writing the software. :-) I might implement 12-term calibration at some point, but it's not high on my list. I'm still not sure it would be very useful. :-) -- Rune / 5Q5R On Mon, 9 Dec 2019 at 09:24, Gabriel Tenma White <OwOwOwOwO123@...> wrote: This might be unintuitive at first but long story short, a T/R VNA *can* |
Re: The T-Check confusion
Kurt,
It appears (in my limited understanding) the poor matching to Z0 of CH1 (port 2) is the biggest reason why the T-check is not performing well on the nanoVNA. Would you agree that using a 10dB attenuator between DUT and CH1 (port 2) would solve most of this? What would be the impact of the remaining non-compensated delay? Should we add to the instructions that, for best nanoVNA 2 port DUT measurements one should seriously consider using a 10dB attenuator between DUT and CH1 (port 2) till new HW becomes available that solves the poor matching of CH1 (port2)? -- NanoVNA Wiki: /g/nanovna-users/wiki/home NanoVNA Files: /g/nanovna-users/files Erik, PD0EK |
Re: Brand new user here
Stylus: Guitar pick works also
toggle quoted message
Show quoted text
STX VHF/UHF Rover On Dec 8, 2019, at 9:49 PM, k9wkj <k9wkjham@...> wrote: |
Re: Brand new user here
Bob Albert
I find metallic works well.
toggle quoted message
Show quoted text
On Sunday, December 8, 2019, 07:49:07 PM PST, k9wkj <k9wkjham@...> wrote:
i use a pen cap or a tooth pick its a resistive screen so anything pointy and non metalic will work even a #2 pencil |
to navigate to use esc to dismiss