Re: How much calibration is enough?
#calibration
On Mon, Apr 27, 2020 at 02:13 AM, Roger Need wrote:
I have made a table to better see what is the result of the measurements from the PDF file. For my understanding here the *precision of the frequency* is important. Calibration MHz Scan MHz Step KHz SWR RL dB Lag KHz 0.05 - 900 0.05 - 900 9000 1.25 19.04 153.041-147.76= 5281 0.05 - 900 140 - 155 150 1.06 29.6 148.400-147.76= 800 140 - 155 140 - 155 150 1.07 29.13 147.600-147.76= 200 140 - 155 146 - 150 40 1.06 29.61 147.800-147.76= 40 146 - 150 146 - 150 40 1.06 30.34 0 *Result* : The frequency lag of maximum RL can not be (much) better, than the measuring step size. So, watch the *Step Size* = scan width / 100. 73, Rudi DL5FA |
Re: OT: looking for testers for a tiny handheld spectrum alayzer
On Mon, Apr 27, 2020 at 01:51 PM, <aleks07111971@...> wrote:
Not sure yet. Target is comparable to one of the nanoVNA family -- NanoVNA Wiki: /g/nanovna-users/wiki/home NanoVNA Files: /g/nanovna-users/files Erik, PD0EK |
Re: OT: looking for testers for a tiny handheld spectrum alayzer
If you really still need testers, I might be able to contribute.
toggle quoted message
Show quoted text
For my day job, I use an HP3585A for phase noise measurements.? Impressive dynamic range on that tank. Also a Rigol for field measurements of RFI in the HF band at the 'other' US radio quiet zone you nave never heard about.? Near Boulder Colorado. I'm doing radio astronomy and HF and LVHF and have several NIST calibrated diode noise sources. If I can help. Terry W0ASP On 4/27/20 8:43 PM, Terry VK5TM wrote:
|
Re: SWR with a NanoVNA
Dave is correct. The VSWR doesn't vary along the transmission line except
for a small variation due to loss of the transmission line. If you do see the VSWR vary, that indicates imbalance on the transmission line. For coax, this means RF is on the outside of the coax shield. 73, Carey, WB4HXE On Mon, Apr 27, 2020 at 4:38 AM Dr. David Kirkby, Kirkby Microwave Ltd < drkirkby@...> wrote: On Fri, 24 Apr 2020 at 15:51, <whdpc113@...> wrote:-- Carey Fisher careyfisher@... |
Re: Up-grading the -F model...
On Mon, Apr 27, 2020 at 03:16 PM, Eric Furness wrote:
I Upgraded to 0.1.1 All worked ok. How do you get to the TDR function? ===================================================== DISPLAY->TRANSFORM From there you have to choose one of the step functions (see various user manuals). Works the same as on on NanoVNA. - Herb |
Re: Up-grading the -F model...
Eric Furness
On 4/27/2020 5:38 AM, June wrote:
v0.1.4 2020-03-26 : This update requires recalibration of the deviceEric WA3UYI -- This email has been checked for viruses by Avast antivirus software. |
Re: NanoVNA firmvare, compiled by DiSlord
#firmware
Only if have problem, if you get correct result, not need do it.
Recalibrate need if threshold value change (possible not correct calibration point near harmonic change) |
Re: NanoVNA firmvare, compiled by DiSlord
#firmware
On Mon, Apr 27, 2020 at 01:18 PM, DiSlord wrote:
Please not use old calibration, recalibrate after update ========================================= DiSlord, I haven't had any issues with 0.8.4.3 or 0.8.4.4 but I do have a question. After updating to each version my calibrations in each slot were still good, when re-checked using my SOLT standards, so I didn't re-calibrate. Is it absolutely necessary to re-calibrate after updating or only if you are experiencing issues? - Herb |
Re: NanoVNA firmvare, compiled by DiSlord
#firmware
DiSlord,
toggle quoted message
Show quoted text
Flashed number 3, calibrated NanoVNA-Saver, measured S21 and 49.85 standard see the pictures. Call it a day for now. Jos Op 27-4-2020 om 22:18 schreef DiSlord: Please not use old calibration, recalibrate after update |
Re: NanoVNA firmvare, compiled by DiSlord
#firmware
i don`t understand
1 variant is last firmware 2 variant - is 0.8.4.3 (Only 8kHz IF change) 3 variant - is old 0.8.4.4 (2x external freq) You not fogot reset self NanoVNA calibration before calibrate in NanoVNA Saver? Try restore last worked firmware |
Re: NanoVNA firmvare, compiled by DiSlord
#firmware
DiSlord,
toggle quoted message
Show quoted text
Flashed number 2,? threshold set to 300MHz,? recalibrated NanoVNA-Saver,? S21 at 0 now,? R+jX? R still not shown, S11 49.85 Ohms standard? way off , Marker 1 an 3 not shown. See pictures. Jos Op 27-4-2020 om 22:15 schreef DiSlord: Check this 3 variants (all of them work good on my H4, try set threshold to 300MHz) |
Re: NanoVNA firmvare, compiled by DiSlord
#firmware
DiS;lord,
toggle quoted message
Show quoted text
Flashed number 1, recalibrated in NanoVNA-Saver, still no good, see picture Jos Op 27-4-2020 om 22:15 schreef DiSlord: Check this 3 variants (all of them work good on my H4, try set threshold to 300MHz) |
Re: NanoVNA firmvare, compiled by DiSlord
#firmware
Please not use old calibration, recalibrate after update
|
Re: NanoVNA firmvare, compiled by DiSlord
#firmware
Check this 3 variants (all of them work good on my H4, try set threshold to 300MHz)
Need recalibrate NanoVNA-H4 96kADC 1.dfu
NanoVNA-H4 96kADC 1.dfu
NanoVNA-H4 96kADC 2.dfu
NanoVNA-H4 96kADC 2.dfu
NanoVNA-H4 96kADC 3.dfu
NanoVNA-H4 96kADC 3.dfu
|
Re: NanoVNA firmvare, compiled by DiSlord
#firmware
DMR
In version v0.8.4.4, the level of CH1 rises after 300 MHz.
|
Re: NanoVNA firmvare, compiled by DiSlord
#firmware
Hi DiSlord,
toggle quoted message
Show quoted text
Loaded your latest version of00000000000000 firmware 0.8.4.4 into my NanoVNA H4 and calibrated in NanoVNA-Saver. Results are not as good as in previous firmware see the pictures attached, sorry. Measuring my 49.85 Ohms standard as 54.xx Ohms , ? when I measure CH1, R? is not shown in R+jX and S21 is not 0. . See also my pictures taken with the previous firmware I had sent. Regards Jos Op 27-4-2020 om 19:58 schreef DiSlord: In 0.8.4.4 i only change IF to 8kHz |
Re: OT: looking for testers for a tiny handheld spectrum alayzer
Erik
I would like to be a beta tester. I have a RIGOL DSA1030 Spectrum Analyzer that goes from 9 KHz to 3000 MHz, a sweep generator that goes from 1 MHz to 1500 MHz, a RIGOL DS1202Z-E Digital Storage Oscilloscope that is 200 MHz dual trace with the software package to analyze various serial comm standards, and a Motorola R2200 Service Monitor that goes up to 1000 MHz. If you have any questions please contact me. 73 Darrell Davis KT4WX |
Re: NanoVNA firmvare, compiled by DiSlord
#firmware
Long read TLV320AIC3204 datasheet (unfortunately it is poorly written)
Revert external freq set 96kHz mode for AIC ADC (this mode have more noise), use internal multiplier as before (i found PLL set error in AIC init, and hope fix it) Not use fractional multiplier for AIC PLL clock (look like less noise) Add delay for SI5351 PLL reset (more stable on band change) Some code optimization and definition fore easy control build PS in CH1 mode noise level drop ~3-5dB PSS i hope now not need change threshold |