Re: Locked, then white screen on NanoVNA-H

 

Tried this and still just a white screen. I guess it's bricked. Thanks for all the responses.
Perry


NanoVNA-H(4)

 

Dear,

I have a problem with the NanoVNA-H(4) and NanoVNA-App

I can connect it on com5 but it does not read the data and mention always a number of points error

Does someone has a trick to solve this problem ?

thank you Jean Marie


Re: Which cable measuring method is more accurate - IYHO

 

Hi, out of curiosity I've made a measurement of my nanovna H4 stock RG316 cable (of length 163mm end to end, incl connectors).
NanoVna App shows 158.4mm, while H4 (Measure->Cable) shows almost identical picture as Dislord's one, but with Z0=9k and length 1.6meter (instead of ~Z0=50ohm as on his picture below.). Both measurements done with open end. H4 flashed with today's v1.2.


Re: spreadsheet for importing snp files, Excel 2010

 

Thank you for doing this for us
Mike


Re: spreadsheet for importing snp files, Excel 2010

 

On 2022-05-30 09:44:-0700, you wrote:

Rich, What is SWR:Linear?
Not sure. I was using the spreadsheet to
1. Confirm that menu items were returning values I expected, so that I could update the manual with meaningful language
2. Explore data relationships.

I provided the spreadsheet in a separate thread so that
1. It would be easier to find, rather than being an offhand reply to a different message
2. Provide a path for those not as familiar with spreadsheets to be able to import data easier, and to see how the complex data is handled by Excel. Many of the formulae are in the user manual posted in the files section, so folks wanting to create their own spreadsheets (for plotting and comparisons, data exploration, etc) can use those equations to process the data once it is in complex form.

Then it's up to users to decide how to use it.

~R~


Re: Nanovnasaver plots

 

On Mon, May 30, 2022 at 12:50 AM, tuckvk3cca wrote:


Thanks, just been reading the late G3TXQ articles. Very useful. He said a one
port measurement is no good. Surely if I terminate the choke with a r ohm
resistor , I can measure R + j X that way and subtract to get (R - r) + j X.
Has anybody tried that?
One port S11 measurement on a NanoVNA-H4 gives good results up to several thousand ohms (see attached 3K ohm example). It is better than using the S21 serial method. The S21 serial method does not work that well on a NanoVNA. The reason is that the Ch0 and CH1 impedances are not 50 ohms across a wide range of frequencies and the NanoVNA does not do a 12 point calibration. There have been several discussions in this group about measuring chokes, baluns and impedance using various methods. Lots of good info if you do a group search.

Roger


Re: spreadsheet for importing snp files, Excel 2010

 

Rich, What is SWR:Linear?

Roger


spreadsheet for importing snp files, Excel 2010

 

I updated this spreadsheet for the convenience of others by rearranging the columns and including instructions in the spreadsheet.

~R~
72/73 de Rich NE1EE
The Dusty Key
On the banks of the Piscataqua


Re: Nanovnasaver plots

 

Thanks, Victor!

73,
Costin, YO8RCD

?n lun., 30 mai 2022 la 12:16, Victor Reijs <pe1atn.victor.reijs@...>
a scris:

For a choke we want to determine high im[edance values and the one port is
not really good in that (low accuracy). See here for some background:

The Series through is better (method c)


All the best,

Victor

Op ma 30 mei 2022 om 09:50 schreef tuckvk3cca <tuckvk3cca@...>:

Thanks, just been reading the late G3TXQ articles. Very useful. He said
a
one port measurement is no good. Surely if I terminate the choke with a
r
ohm resistor , I can measure R + j X that way and subtract to get (R -
r)
+ j X. Has anybody tried that?










Re: Question S21 measurement relay

 

Alright, thank you!

Best
Fabio


Re: Nanovnasaver plots

 

For a choke we want to determine high im[edance values and the one port is
not really good in that (low accuracy). See here for some background:

The Series through is better (method c)


All the best,

Victor

Op ma 30 mei 2022 om 09:50 schreef tuckvk3cca <tuckvk3cca@...>:

Thanks, just been reading the late G3TXQ articles. Very useful. He said a
one port measurement is no good. Surely if I terminate the choke with a r
ohm resistor , I can measure R + j X that way and subtract to get (R - r)
+ j X. Has anybody tried that?






Re: NanoVNA Saver 0.4.0 MacOS #nanovna-saver

 

Solved : just change line thickness and point size before closing window


NanoVNA Saver 0.4.0 MacOS #nanovna-saver

 

Hi all,

I don't know if this has been posted before, but I'll give it a try.
I use the software on my Mac and all seems to run. But when I go to 'Display Setup' and check 'Show lines' the software crashes. The only way to recover and make it possible to start is to edit or delete the NanoVNASaver.ini file. Is there a solution to display lines in stead of dots?

Many thanks,
Alex, PE1EVX


Re: Nanovnasaver plots

 

Thanks, just been reading the late G3TXQ articles. Very useful. He said a one port measurement is no good. Surely if I terminate the choke with a r ohm resistor , I can measure R + j X that way and subtract to get (R - r) + j X. Has anybody tried that?


Re: Nanovnasaver plots

 

On Sun, May 29, 2022 at 12:50 AM, tuckvk3cca wrote:


Great thanks. Any body has a spreadsheet then to convert S21 gain and phase
data into impedances?
There are two methods to Convert S21 to Z. The series method is used for "high impedances" and the "shunt method" for low impedances. These are built into the latest firmware for H/H4 and in NanoVNA Saver and NanoVNA app. You need a very good test jig for S21 series or the phase angle will have errors and the impedance measured will be way off. Also the H/H4 do not have precise 50 ohm source and load impedance so you get errors from that as well.

But to directly answer your question here is is a link to a Web site with a spreadsheet. The link is in the last paragraph.


Here is a link to another one....
/g/nanovna-users/message/17807

Roger


nVNA-H4 v4.3_MS, firmware 1.1, User Manual 0100

 

I posted an updated version of this doc, which is the same content as the older doc, but because it contains more operational information, I renamed it to a user manual. I thought that I had sent a message to the group when it posted, but I didn't see the message.

We now have a newer firmware version, so I'll see if I can find the time to update...

~R~
72/73 de Rich NE1EE
The Dusty Key
On the banks of the Piscataqua


File Notifications #file-notice

Group Notification
 

The following files and folders have been deleted from the Files area of the [email protected] group.

* /user manual and menu maps/nVNA-H4 v4.3_MS, firmware 1.1, Menu Items & Descriptions 0110..zip

*By:* Rich NE1EE <TheDustyKey@...>



---

The following files and folders have been uploaded to the Files area of the [email protected] group.

* /user manual and menu maps/nVNA-H4 v4.3_MS, firmware 1.1, User Manual 0100.zip ( /g/nanovna-users/files/user%20manual%20and%20menu%20maps/nVNA-H4%20v4.3_MS,%20firmware%201.1,%20User%20Manual%200100.zip )

*By:* Rich NE1EE <TheDustyKey@...>

*Description:*
nVNA-H4 v4.3_MS, firmware 1.1, User Manual 0100 latest version. includes more operational notes. Word and PDF.


Re: Locked, then white screen on NanoVNA-H

 

Yes, disconnect all power for a while. The ultimate and first try to fix ANY and EVERY computer that I know of. Even my wife has learned that.

Dave K8WPE

David J. Wilcox¡¯s iPad

On May 28, 2022, at 1:40 PM, Roger Need via groups.io <sailtamarack@...> wrote:

?On Fri, May 27, 2022 at 04:15 AM, Perry Westberry wrote:


Yes, I adjusted the whip. No transmitter use. Can it be fixed?
Perry
One thing you might try is to charge the unit overnight. Then open the case and then disconnect the battery by taking the connector off the board. Wait a minute or so and then re-connect the battery. This may help if the CPU chip is in some unknown state....

Roger





Re: Locked, then white screen on NanoVNA-H

 

Thanks, Roger. I'll give that a try.
Perry

On Sat, May 28, 2022 at 1:40 PM Roger Need via groups.io <sailtamarack=
[email protected]> wrote:

On Fri, May 27, 2022 at 04:15 AM, Perry Westberry wrote:


Yes, I adjusted the whip. No transmitter use. Can it be fixed?
Perry
One thing you might try is to charge the unit overnight. Then open the
case and then disconnect the battery by taking the connector off the
board. Wait a minute or so and then re-connect the battery. This may
help if the CPU chip is in some unknown state....

Roger






Re: Nanovnasaver plots

 

On 2022-05-29 00:50:-0700, you wrote:

Great thanks. Any body has a spreadsheet then to convert S21 gain and phase data into impedances?
You might find this useful...I did it with real data from my -H4 and my EF 40 m wire. You can make a copy of this spreadsheet so that you have a backup.
Then open it,
and start a new spreadsheet.
Select Data > From Text
In the File Selector mask drop box, select all files *.*
Browse to your s1p / s2p files
import one
start at line ? depending on what line data starts in your snp file.
should read right in.
----- either
copy your data in cols to this spreadsheet, noting that it has a col inserted where freq is converted to MHz
if your data is shorter (fewer rows) delete the extra rows
-----or
copy the formulas from this spreadsheet to your spreadsheet, paying attention to changing cols

It's just some thing I tossed together to look at some data but
S11, SWR, Linear, SWR:Linear, Z, |Z|, Q
are all good calculations

~R~
72/73 de Rich NE1EE
The Dusty Key
On the banks of the Piscataqua