Re: calibration skewed


 

Hi Paul
Have you tested you load for being 50 ohm. Mine went bust pretty quikly so try a different load
Kind regards
Kurt
-----Oprindelig meddelelse-----
Fra: [email protected] <[email protected]> P? vegne af paul larner
Sendt: 3. december 2019 08:58
Til: [email protected]
Emne: [nanovna-users] calibration skewed

I set the sweep from 1 to 30mhz then do the cal save in pos 0 then restart it,in the smith plot the marker moves all the way to the left like it should with cho shorted but with it open the marker sits about 3/4 of an inch from thr right hand side not fully to the right,any idea why this is?,im using the cal kit that came with it.