Thank you! That is what I was planning on doing, unfortunately no access to
lab grade loads. I'll have to do the best I can...
Tim K4SHF
On Sat, Dec 18, 2021 at 7:46 PM Donald S Brant Jr <dsbrantjr@...>
wrote:
I would use a known-good set of calibration standards to calibrate the
VNA, then measure the other standards to judge their quality.
On a related subject, measuring a load or other calibration standard by
putting it back after using it to calibrate the system "to check the
calibration" will only measure your connection repeatability. The
standard will always measure as perfect (or nearly so) because the system
has essentially been told via the calibration standards definitions "this
is a perfect 50¦¸ termination" even if it is not.
73, Don N2VGU