Back in business
JerryM
Okay, I have been having trouble achieving a suitable vacuum. Today i replaced the o-ring seal on my diffusion pump. While I had it out I found the nameplate that said "use 50cc of DP oil. When I poured it out (much spilling) I discovered that I had way too much. Probably 100cc.
After adjusting the quantity the pump worked like new. I got to 10X10-5, without using the cold trap. I am sure that with the cold trap I will do much better. On to coat some mirrors. Jerry |
JerryM
oops,
toggle quoted message
Show quoted text
Still had a problem where it wouldn't pump below 100 microns. Upon examination I found that air was trapped in the high current feed throughs (yes both of them) Hopefully cutting a relief channel will solve this problem. I am pumping as I write. Jerry --- In VacuumX@..., "JerryM" <gandamiller@...> wrote:
|
JerryM
--- In VacuumX@..., "JerryM" <gandamiller@...> wrote:
As I said on to coat mirrors. I just uploaded a photo of the setup in my bell jar. The mirrors have a small radius and would not coat on the edges. That is why there is a small motor inside to turn the mirror during coating. this setup is very successful. Jerry |